SPIRou Specifications in a Nutshell
|
Last update: April 2017.
The numbers below show the main instrument technical specifications derived from the science requirements.
System |
Throughput | > 10% |
Thermal background at 2.35 microns | < 50 ph/s/A |
Sensitivity of the guiding | up to H~14 |
Spectrograph |
Wavelength Domain | 0.98-2.35 microns |
Spectral Resolution | > 70,000 |
Pixel size | < 2.3 km/s |
Velocimeter |
Radial-velocity accuracy | 1 m/s (internal) |
Method | simultaneous spectral calibration |
Detector |
Type | H4RG |
Dimensions | 4k x 4k |
Pixel size | 15 microns |
Quantum Efficiency | > 70% |
Cosmetics | < 1% bad pixels |
Readout mode | Ramp fitting |
Dark current | < 0.05 e-/s |
Readout Noise (CDS) | < 15 e- |
|
|